CVE-2017-9657
Description
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Weakness
- CWE-460: Improper Cleanup on Thrown Exception
Timeline
- 2018-04-30Published by the CNA15:00 UTC
Affected products
Products listed as vulnerable. Upgrade to the version under "Fixed in", or a later one.
| Product | Vulnerable versions | Fixed in |
|---|---|---|
| philips intellivue_mx40_firmware |
References
- https://www.usa.philips.com/healthcare/about/customer-support/product-security
- http://www.securityfocus.com/bid/100813
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01
Get more CVE data from the API.
curl 'https://api.contrastcyber.com/cve_lookup/CVE-2017-9657'curl 'https://api.contrastcyber.com/cve_risk_score/CVE-2017-9657'curl 'https://api.contrastcyber.com/cve_exploit/CVE-2017-9657'curl 'https://api.contrastcyber.com/cve_products/CVE-2017-9657'
Sources
- Vulnerability data from the National Vulnerability Database (NIST) and the CVE Program. CVE records: Copyright © 1999-2026, The MITRE Corporation, used under the CVE Program Terms of Use. CVE® is a registered trademark of The MITRE Corporation.
- CWE definitions: Copyright © 2006-2026, The MITRE Corporation, used under the CWE Terms of Use. CWE is a trademark of The MITRE Corporation.
- OWASP Top 10 categories by the OWASP Foundation (CC BY 3.0).
- EPSS scores by FIRST.org EPSS, used with attribution.
- KEV catalog and SSVC decisions by CISA.
- Advisory data from the GitHub Advisory Database through OSV.dev (CC BY 4.0).
- Exploit metadata from the Exploit Database (GPL-2.0-only, metadata only).
- Nuclei template names from nuclei-templates (MIT).