CVE-2017-9658

Description

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Severity
MEDIUM
CVSS v3
6.5
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H · source nvd
Exploit chance (EPSS)
0.810%
chance of exploitation in the next 30 days
higher than 57.5% of scored CVEs · 2026-11-17
CISA KEV
Not listed
Public exploit code
None public
Exploit references (NVD and CNA): 0
Metasploit modules: 0
Nuclei templates: 0

Weakness

Timeline

  1. 2018-04-30Published by the CNA15:00 UTC

The timeline shows what already happened. To be told when this CVE changes next, watch it from the API.

Affected products

Products listed as vulnerable. Upgrade to the version under "Fixed in", or a later one.

ProductVulnerable versionsFixed in
philips intellivue_mx40_firmware

Products and versions as NVD and the vendor that published the CVE list them.

References

Get more CVE data from the API.

Free without a key or signup. Pro adds monitoring and higher limits: pricing.

Sources