CVE-2018-4846

Description

A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

Severity
CRITICAL
CVSS v3
9.8
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H · source nvd
Exploit chance (EPSS)
1.734%
chance of exploitation in the next 30 days
higher than 78.1% of scored CVEs · 2026-11-17
CISA KEV
Not listed
Public exploit code
None public
Exploit references (NVD and CNA): 0
Metasploit modules: 0
Nuclei templates: 0

Weakness

Timeline

  1. 2018-06-26Published by the CNA18:00 UTC

The timeline shows what already happened. To be told when this CVE changes next, watch it from the API.

Affected products

Products listed as vulnerable. Upgrade to the version under "Fixed in", or a later one.

ProductVulnerable versionsFixed in
siemens rapidlab_1200_firmwarebefore 3.3fixed in 3.33.3
siemens rapidpoint_400_firmware
siemens rapidpoint_500_firmwarethrough 2.3no fix listedfrom 3.0no fix listednot listednot listed

Products and versions as NVD and the vendor that published the CVE list them.