CVE-2019-10964
Description
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Severity
HIGH
CVSS v3
7.1
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H · source cna
Exploit chance (EPSS)
1.234%
chance of exploitation in the next 30 days
higher than 69.5% of scored CVEs · 2026-11-17
CISA KEV
Not listed
Public exploit code
None public
Exploit references (NVD and CNA): 0
Metasploit modules: 0
Nuclei templates: 0
Weakness
- CWE-284: Improper Access Control
Timeline
- 2019-06-28Published by the CNA20:58 UTC
Affected products
Products listed as vulnerable. Upgrade to the version under "Fixed in", or a later one.
| Product | Vulnerable versions | Fixed in |
|---|---|---|
| medtronic minimed_508_firmware | ||
| medtronic minimed_paradigm_511_firmware | ||
| medtronic minimed_paradigm_512_firmware | ||
| medtronic minimed_paradigm_515_firmware | ||
| medtronic minimed_paradigm_522_firmware | ||
| medtronic minimed_paradigm_522k_firmware | ||
| medtronic minimed_paradigm_523_firmware | through 2.4ano fix listed | not listed |
| medtronic minimed_paradigm_523k_firmware | through 2.4ano fix listed | not listed |
| medtronic minimed_paradigm_712_firmware | ||
| medtronic minimed_paradigm_712e_firmware | ||
| medtronic minimed_paradigm_715_firmware | ||
| medtronic minimed_paradigm_722_firmware | ||
| medtronic minimed_paradigm_722k_firmware | ||
| medtronic minimed_paradigm_723_firmware | through 2.4ano fix listed | not listed |
| medtronic minimed_paradigm_723k_firmware | through 2.4ano fix listed | not listed |
| medtronic minimed_paradigm_veo_554_firmware | through 2.6ano fix listed | not listed |
| medtronic minimed_paradigm_veo_554cm_firmware | through 2.7ano fix listed | not listed |
| medtronic minimed_paradigm_veo_754_firmware | through 2.6ano fix listed | not listed |
| medtronic minimed_paradigm_veo_754cm_firmware |
References
- http://www.securityfocus.com/bid/108926
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-508-paradigm.html
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-178-01
- https://www.us-cert.gov/ics/advisories/icsma-19-178-01
Get more CVE data from the API.
curl 'https://api.contrastcyber.com/cve_lookup/CVE-2019-10964'curl 'https://api.contrastcyber.com/cve_risk_score/CVE-2019-10964'curl 'https://api.contrastcyber.com/cve_exploit/CVE-2019-10964'curl 'https://api.contrastcyber.com/cve_products/CVE-2019-10964'
Sources
- Vulnerability data from the National Vulnerability Database (NIST) and the CVE Program. CVE records: Copyright © 1999-2026, The MITRE Corporation, used under the CVE Program Terms of Use. CVE® is a registered trademark of The MITRE Corporation.
- CWE definitions: Copyright © 2006-2026, The MITRE Corporation, used under the CWE Terms of Use. CWE is a trademark of The MITRE Corporation.
- OWASP Top 10 categories by the OWASP Foundation (CC BY 3.0).
- EPSS scores by FIRST.org EPSS, used with attribution.
- KEV catalog and SSVC decisions by CISA.
- Advisory data from the GitHub Advisory Database through OSV.dev (CC BY 4.0).
- Exploit metadata from the Exploit Database (GPL-2.0-only, metadata only).
- Nuclei template names from nuclei-templates (MIT).