CVE-2020-12048

Description

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

Severity
HIGH
CVSS v3
7.5
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N · source nvd
Exploit chance (EPSS)
0.452%
chance of exploitation in the next 30 days
higher than 39.8% of scored CVEs · 2026-11-17
CISA KEV
Not listed
Public exploit code
None public
Exploit references (NVD and CNA): 0
Metasploit modules: 0
Nuclei templates: 0

Weakness

Timeline

  1. 2020-06-29Published by the CNA13:48 UTC

The timeline shows what already happened. To be told when this CVE changes next, watch it from the API.

Affected products

Products listed as vulnerable. Upgrade to the version under "Fixed in", or a later one.

ProductVulnerable versionsFixed in
baxter phoenix_x36_firmware3.36no fix listed3.40no fix listednot listednot listed

Products and versions as NVD and the vendor that published the CVE list them.