CVE-2022-29953

Description

The Bently Nevada 3700 series of condition monitoring equipment through 2022-04-29 has a maintenance interface on port 4001/TCP with undocumented, hardcoded credentials. An attacker capable of connecting to this interface can thus trivially take over its functionality.

Severity
CRITICAL
CVSS v3
9.8
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H · source nvd
Exploit chance (EPSS)
1.067%
chance of exploitation in the next 30 days
higher than 65.3% of scored CVEs · 2026-11-17
CISA KEV
Not listed
Public exploit code
None public
Exploit references (NVD and CNA): 0
Metasploit modules: 0
Nuclei templates: 0

Weakness

Timeline

  1. 2022-07-26Published by the CNA21:42 UTC

The timeline shows what already happened. To be told when this CVE changes next, watch it from the API.

Affected products

Products listed as vulnerable. Upgrade to the version under "Fixed in", or a later one.

ProductVulnerable versionsFixed in
bakerhughes bently_nevada_3701/40_firmwarebefore 4.1fixed in 4.14.1
bakerhughes bently_nevada_3701/44_firmwarebefore 4.1fixed in 4.14.1
bakerhughes bently_nevada_3701/46_firmwarebefore 4.1fixed in 4.14.1
bakerhughes bently_nevada_60m100_firmware

Products and versions as NVD and the vendor that published the CVE list them.